evidenceevidencedocumentFederal Criminal TrialOctober 2013 Theranos emails on HCG test precision and allowable errorAdam Rosendorff explained that a test needs to be especially precise near a threshold that doctors use to make a clinical decision. At 50 milliunits per milliliter, variation in repeat measurements was 21 percent, above the 15 percent limit for total error.
Adam Rosendorff explained that a test needs to be especially precise near a threshold that doctors use to make a clinical decision. At 50 milliunits per milliliter, variation in repeat measurements was 21 percent, above the 15 percent limit for total error.